Sign In | Join Free | My ecer.co.uk |
|
Brand Name : KACISE
Model Number : KGD-411
Certification : CE
Place of Origin : CHINA
MOQ : 1PCS
Payment Terms : L/C,D/A,D/P,T/T,Western Union,MoneyGram
Supply Ability : 1000 Piece/Pieces per Week negotiable
Delivery Time : 5-8 work days
Packaging Details : each unit has individual box and all boxes are packed in standard packages or customers requests available
Detection Principle : Non - Dispersive Infrared Technology (NDIR)
Data Output : CO₂ Concentration (mmHg), End - Tidal CO₂, Inhaled CO₂, Respiratory Rate
Detection Range : 0 - 150mmHg
Resolution : 0.1mmHg
Respiratory Rate Range : 0 - 150BPM
Respiratory Rate Accuracy : ±1 breath
Sidestream End - Tidal Carbon Dioxide Sensor KGD - 411 for Measuring CO₂ Concentration
Product Introduction
The KACISE sidestream end - tidal carbon dioxide sensor is a CO₂ sensor module designed based on the principle of infrared absorption spectroscopy, used for measuring CO₂ concentration. It can be applied in devices such as monitors, nutritional metabolism analyzers, and exercise cardiopulmonary function testers. It has the advantages of high precision, good temperature characteristics, low power consumption, fast response time, compact appearance, and flexible integrated installation.
Product Features
Technical Parameters
Items | Details |
Detection Principle | Non - Dispersive Infrared Technology (NDIR) |
Data Output | CO₂ Concentration (mmHg), End - Tidal CO₂, Inhaled CO₂, Respiratory Rate |
Detection Range | 0 - 150mmHg |
Resolution | 0.1mmHg |
Detection Accuracy | 0 - 40mmHg: ±2mmHg |
41 - 70mmHg: True Value ±5% | |
71 - 100mmHg: True Value ±8% | |
100 - 150mmHg: True Value ±10% | |
Respiratory Rate Range | 0 - 150BPM |
Respiratory Rate Accuracy | ±1 breath |
Stability | Short - term Drift: Not Exceeding 0.8mmHg (4H) |
Long - term Drift: Maintaining Accuracy Specifications within 120H | |
Input Voltage | 5.00V (±5%) |
Average Current | <100mA |
Peak Current | <200mA |
Initialization Time | Display within 15s, Full Specification Accuracy within 2min (25°C) |
Compensation | Air Pressure: 400mmHg - 850mmHg |
Operating Temperature and Humidity | Operation: 0°C - 55°C, 10 - 90% RH (Non - condensing) |
Storage: -40°C - 70°C, <90% RH (Non - condensing) | |
Data Communication | TTL |
Lifetime | ≥5 Years |
Dimensions | 57.5mm * 23.7mm * 24mm |
![]() |
Sidestream End - Tidal Carbon Dioxide Sensor KGD - 411 for Measuring CO₂ Concentration Images |